GE Dash 4000


Certified Refurbished Medical Equipment!

GE Dash 4000 Features
The Dash 4000 Pro is the most advanced configured monitor available for clinicians who want to improve the quality of their care. Dash monitors feature gold standard Marquette ECG, DINAMAP, NIBP, Masimo SET Pulse Oximetry and other leading patient-safety enhancements in a rugged, ergonomic design light and durable enough to handle even air transport. Modular enhancements and a large 10″ color display make the GE Dash 4000 a good fit for the ED, surgical center and even the ICU.

$3,988.67

GE Dash 4000 Features
The Dash 4000 Pro is the most advanced configured monitor available for clinicians who want to improve the quality of their care. Dash monitors feature gold standard Marquette ECG, DINAMAP, NIBP, Masimo SET Pulse Oximetry and other leading patient-safety enhancements in a rugged, ergonomic design light and durable enough to handle even air transport. Modular enhancements and a large 10″ color display make the GE Dash 4000 a good fit for the ED, surgical center and even the ICU.

No monitor on the market offers more life-saving assessment tools like 12-lead 12SL, EK-Pro arrhythmia, ST-segment analysis, A-Fib detection and IntelliRate alarm reduction for less cost. All GE bedside monitors feature a common Trim Knob user interface and intuitive menu structures, so every clinician can be skilled using the bedside monitor throughout your continuum of care.

Unmatched adaptability at the bedside.

Modular flexibility
Revolutionary chest pain care
Early intervention in the NICU
Gold standard NIBP accuracy
Options: Agent Monitoring
GE Dash 4000 Specifications

Dimensions
Height: 27.4 cm. Depth: 24.3 cm. Width: 29.3 cm.
Weight: 5.7 kg
Display
Size: 26 cm (diagonal).
Type: active-matrix color TFT.
Resolution: 640 by 480 pixels.
Number of traces: 6 maximum.
Number of seconds/trace: 4.9 at 25 mm/sec.
Sweep speed: 6.25, 12.5, 25 mm/sec (with erase bar).
Information window: displays non-real-time information without obstructing the display of real-time information.
Display organization: prioritized by parameter.

Controls
Trim Knob® control.
Five hard keys: silence alarm, graph go/stop, function (zero all) and power on/off.

Alarms
Categories: patient status and system status.
Priority: 4 levels – crisis, warning, advisory, and message.
Notification: audible and visual.
Setting: default and individual.
Silencing: 1 minute, current alarm only.
Pause: 5 minutes in adult ICU mode, 3 minutes in neonatal ICU mode, and 5,15, or permanent pause in OR mode.
Volume: default 70%, 70 dB measured at 1 meter.

ECG
Standard leads available: I, II, III, V, aVR, aVL, and aVF. 10
lead wire cable: I, II, III, V, aVR, aVF, V2, V3, V4, V5, V6.
Leads analyzed simultaneously: I, II, III, and V (multi-lead mode).
Lead fail: identifies failed lead.
Alarms: user-selectable upper and lower heart rate limits.

Input Specifications
Voltage range: ±0.5 mV to ±5 mV.
Signal width: 40 ms to 120 ms (Q to S).
Heart rate range: 30 to 300 bpm.
Input impedance: common mode > 10 Mat 50/60 Hz, differential > 2.5 Mfrom dc to 60 Hz.
Common mode rejection: 90 dB minimum at 50 or 60 Hz

Respiration
Measurement technique: impedance variation detection.
Range respiration rate: 0 – 200 breaths per minute.
Base impedance: 100 – 1000 Ωat 52.6 kHz.
Detection sensitivity: 0.4 to 10 Ωvariation.
Waveform display bandwidth: 0.1 to 1.8 Hz (-3 dB).
Alarms: user-selectable upper and lower respiration limits, and user-selectable apnea limit.

Temperature
Number of channels: 2.
Input specifications – probe type: YSI series 400 or 700,
temperature range: 0°C to 45°C (32°F to 113°F),
resolution: ±0.1°C.
Output specifications – parameters displayed: T1, T2,
accuracy: ±0.1°C for YSI series 4000 probes, ±0.3°C for YSI series 7000 probes, independent of source.
Alarms: user-selectable upper and lower limits for T1 and T2.
Invasive Blood Pressure
Number of channels: 2 (optional).
Transducer sites: arterial, femoral artery, pulmonary arterial, central venous, right atrial, left atrial, intracranial, and special.
Transducer requirements: 5 Vdc±0.1% excitation voltage.
Transducer output: 5 μV/V/mmHg. Input specifications – range: -25 mmHg to 300 mmHg, offset: ±150 mmHg.

Output specifications – frequency response: dc to 50 Hz (-0/+2 Hz).
Zero balance range: ±150mmHg.
Zero balance accuracy: ±1mmHg.
Zero balance drift: ±1mmHg over 24 hours.
Accuracy: ±2% or ±1mmHg, whichever is greater.
Alarms: user-selectable upper and lower limits for systolic, diastolic, and mean pressures.

NIBP
Measurement technique: oscillometric.
Displayed parameters; systolic, diastolic, and mean pressures.
Systolic: 30-275 mmHg adult, 30-235 mmHg pediatric, and 30-135 mmHg neonate.
Diastolic: 10-220 mmHg adult, 10-220 mmHg pediatric, and 10-110 mmHg neonate.
MAP: 20-260 mmHg adult, 20-260 mmHg pediatric, and 20-125 mmHg neonate.
Pulse rate, as displayed in tabular trends: 30-200 bpm adult, 30-200 bpm pediatric, and 30-200 bpm neonate

Pulse Oximetry
Parameters monitored: arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR).
SpO2range: 50 –100%. PPR range: 30-300 beats per minute.
SpO2 accuracy: ±2% (70-100%) and ±3% (50- 69%).
PPR accuracy: ±3 beats per minute.
Alarms: user-selectable upper and lower limits for SpO2 and PPR.

C02
The information displayed: inspired and expired CO2 concentrations in %, mmHg, or a, respiratory rate, and continuous CO2 waveform.
Measurement ranges: PiCO2/FiCO2: 0 to 100 mmHg/0to13%. Pe CO2/ Fe CO2: 0 to 100 mmHg/0 to 13%;
RR: 0 to 120 breaths/min.
Accuracy: CO2: 5% of reading or ±2 mmHg, whichever is greater.
Display resolution: 1 mmHg. Rise time: less than 60 msec.
Respiration rate accuracy: ± breath/min

Categories: ,

Based on 0 reviews

0.0 overall
0
0
0
0
0

Be the first to review “GE Dash 4000”