Philips Respironics V680 Features
The Philips Respironics V680 is a critical care ventilator that combined a single-limb NIV with a fully featured dual-limb system to deliver life support for critically ill patients in the ICU. The Respironics V680 can easily and quickly switch from invasive to non-invasive ventilation. This Philips ventilator provides continuous measurements of dynamic resistance, compliance, elastance, and plateau pressure in both single and dual-limb invasive and non-invasive modes.
Combines the proven performance of the Philips V60 non-invasive ventilator with an invasive ventilator for ICU patients.
Same look and feel as the Respironics V60 Ventilator.
Built-in apnea backup for both invasive and non-invasive spontaneous modes.
4-hour battery backup.
Uses Auto-Trak technology to automatically adapt to your patient’s breathing pattern.
Supports adult and paediatric patients.
Philips Respironics V680 Specifications
Dimensions
Height: 13.5” (34.2 cm)
Width: 18” (45.7 cm)
Depth: 17.1” (43.4 cm)
Weight: 27 lbs (12.3 kg)
Patient Types
Adult: >20 kg
Paediatric: 5 to 20 kg
Pneumatics
High-pressure oxygen supply: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to 87 psig
Flow: 175 SLPM
Air Supply: Integrated centrifugal-flow compressor
Exhalation cartridge (eSYS)
Flow sensor: Exhaled gas flow accuracy: +/- (0.1 SLPM + 5% of re adding)
Diaphragm/seat area: 6.6 cm2
Diaphragm/seat diameter: 29 mm
Oxygen Sensor
Accuracy: +/-5% (calibrated)
T90 Response: 50 sec for VT = 50 mL; 21 sec for VT = 1,000 mL
Ventilation Modes
CPAP (Continuous Positive Airway Pressure)
S/T (Spontaneous with Timed Backup)
PCV (Pressure Control Ventilation)
Apnea mode (available in CPAP)
AVAPS+ (Average Volume Assured Pressure Support)
PPV (Proportional Pressure Ventilation)
A/C-VCV (Assist/Control-Volume Control Ventilation)
A/C-PCV (Assist/Control-Pressure Control Ventilation)
SIMV-VCV (Synchronized Intermittent Mandatory Ventilation Volume Control Ventilation)
SIMV-PCV (Synchronized Intermittent Mandatory Ventilation Pressure Control Ventilation)
PSV (Pressure Support Ventilation)
PRVC (Pressure Regulated Volume Control)
Apnea mode (available in SIMV and PSV)
Waveforms Window
Pressure Waveform: 0 to 70 cmH2O
Flow Waveform: -240 to 240 L/min BTPS
Volume Waveform: 0 to 3,500 mL BTPS
F/V (flow/volume) Loop
Flow: +/-10 to +/-240 L/min
Volume: 50 to 3,500 mL
P/V (pressure/volume) Loop
Pressure: Above zero: 10 to 80 cmH2O; Below zero: 0 to -15 cmH2O
Volume: 50 to 3,500 mL
Single-Limb Settings
CPAP: 4 to 25 cmH2O
EPAP: 4 to 25 cmH2O
IPAP: 4 to 40 cmH2O
I-Time (Inspiratory Time): 0.30 to 3.00 secs
Ramp Time: Off, 5 to 45 min
Rate (Respiratory Rate): 4 to 80 BPM
Rise (Rise Time): 1 to 5
Triggering and Cycling: Auto-adaptive (Auto-Trak+)
O2 % (Oxygen Percent): 21 to 100%
Apnea Mode (available in CPAP mode): Allows apnea mode settings independent of primary mode
C-Flex: Off, 1 to 3
Max ∆P/min (AVAPS Max P change/min): 1.0 to 5.0 cmH2O/min
Max P (AVAPS max IPAP): 6 to 40 cmH2O
Min P (AVAPS min IPAP): 5 to 30 cmH2O
VT (AVAPS Target Tidal Volume): 50 to 2,000 mL BTPS
Max E: 0 to 100 cmH2O/L
Max R: 0 to 50 cmH2O/L/s
PPV%: 0 to 100%
Max P (PPV max pressure limit): 5 to 40 cmH2O
Max V (PPV max volume limit): 200 to 3,500 mL
Dual-Limb Settings
VT (Tidal Volume in VCV and PRVC modes): 50 to 2,000 mL BTPS
Rate (Respiratory Rate): 1 to 80 BPM
I-Time (Inspiratory Time): 0.30 to 5.00 secs
Rise (Rise Time): 1 to 5
I-Trig (Flow Trigger): 0.5 to 20 L/min, Off
E-Cycle (% of Peak Flow): 10 to 80%
Flow Pattern: Square, Descending Ramp
PC (Pressure Control Target above PEEP): 1 to 65 cmH2O
PS (Pressure Above PEEP): Off, 2 to 65 cmH2O
PEEP (End Expiratory Pressure): 0 to 40 cmH2O
Sigh (1.5 times Tidal Volume): On, Off
O2 % (Oxygen Percent): 21 to 100%
Max P (PRVC maximum pressure limit): 3 to 65 cmH2O
Max V (PRVC maximum volume limit): 55 to 2,500 mL
Min P (PRVC minimum pressure limit): 2 to 64 cmH2O
Apnea Mode (available in PSV and SIMV modes): Allows apnea mode settings.