Drager Fabius MRI


Certified Refurbished Medical Equipment!

Drager Fabius MRI Features
The Drager Fabius MRI is an anaesthesia machine that is compatible with an MRI Machines. I can be used with MRI Systems of 1.5 and 3 Tesla. The Fabius MRI offers a variety of ventilation modes including volume and pressure. The Drager MRI Anaesthesia Machine comes with an E-vent piston-controlled ventilator – this can deliver a maximum peak flow of ventilation and does not require a drive gas. The anaesthesia Machine has a large, high-visibility colour screen along with an integrated warning and alarm LED system.

$25,270.00

Drager Fabius MRI Features
The Drager Fabius MRI is an anaesthesia machine that is compatible with an MRI Machines. I can be used with MRI Systems of 1.5 and 3 Tesla. The Fabius MRI offers a variety of ventilation modes including volume and pressure. The Drager MRI Anaesthesia Machine comes with an E-vent piston-controlled ventilator – this can deliver a maximum peak flow of ventilation and does not require a drive gas. The anaesthesia Machine has a large, high-visibility colour screen along with an integrated warning and alarm LED system.

High visibility colour screen.
Integrated Warning and alarm LEDs.
2 vaporizer mount.
Pressure gauges for pin index gas cylinders.
Dräger Sob CLIC absorber.
Anesthetic Gas Scavenging System.
3 Large drawers, 1 with lock.
Central brake.
Compact Breathing System.
E-vent piston Ventilator.
Dräger Fabius MRI Specifications
Base Unit
Trolley Version (Cart) with COSY: Dimensions (W × H × D): approximately 39 × 55 × 35.5 in. (99 × 140 × 90 cm)
Fabius MRI Trolley (with COSY) Weight and load without supplementary cylinders and vaporizers: 365 lbs. (165.8 kg)
Power and Battery Backup
Power Input: 100 to 240 VAC, 50 / 60 Hz, 70 VA, including additional power outlets.
Operation time with fully charged batteries: > 45 min.
Anesthesia Gas Supply Module
Range of fresh gas flow indicators: 0.0 to 12.0 L/min Total fresh gas flowmeter 0 to 10 L/min.
O2 flush: at 87 psi (6 bar): max 75 L/min; at 41 psi (2.8 bar): min. 25 L/min.
Vaporizer: 2 position Dräger mount (Interlock 2 – System): Dräger Isoflurane Vapor 2000, Dräger Sevoflurane Vapor 2000, Dräger Halothane Vapor 2000, Dräger Enflurane Vapor 2000.
Ventilator Operating Specifications
Ventilator E-vent: Electronically controlled, electrically driven.
Operating modes: Volume Controlled Ventilation, Pressure Controlled Ventilation, Pressure Support, SIMV/PS, Manual Ventilation, Spontaneous Breathing.
Control Input Ranges
Breathing Frequency (rate): 4 to 60 bpm.
Positive End Expiratory Pressure (PEEP): 0 to 20 cmH2O (hPa).
Inspiration/expiration ratio (Ti:Te): 4:1 to 1:4.
Pressure limiting (Pmax): 15 to 70 cmH2O (hPa).
Tidal Volume (Vt): 20 to 1400 mL in Volume Control, 20 to 1100 mL in SIMV/PS.
Inspiration pause (Tip:Ti): 0 to 50 %.
SIMV Inspiratory time: 0.3 – 4.0 sec.
Inspiratory pressure (Pinsp): PEEP + 5 to 65 cmH2O (hPa).
Inspiratory Flow (InspFlow): 10 to 75 L/min in Volume and Pressure Control modes, 10 to 85 L/min in Pressure Support and SIMV/PS modes.
Pressure Support Level (PPS): PEEP + 3 to 20 cmH2O (hPa) Min. frequency for apnoe-ventilation (Freq. Min.) 3 to 20 bpm and “OFF”.

Categories: ,